MS400 Multiparameter Simulator is a portable device, which can provide comprehensive physiological waveform to content general test and train requirement.
It can check the monitor rapidly near the bed, also can test the quality warranty of equipment, monitor and remote therapy equipment, At the same time,
it can be used as the training of distinguishing from arrhythmia waveforms to the nurses.
3.5-inch color TFT touch-screen, 320*240 pixels, friendly operation interface.
Display the output waveforms synchronously, which is convenient for user to compare with the test equipment.
It integrates the functions of SD socket interface, user can output the custom waveform by this device.
Built-in templates list to provide the waveform that is expressed by each preset sequence number,which is convenient for searching.
Two operation interfaces:Chinese and English.
Operate with keystroke and touch-screen expediently and rapidly.
Built-in 3.7V/4000mAh(Rechargeable alkaline battery).
Optical external power: AC 220V, 50Hz or AC 120V, 60Hz
/Arrhythmia
Conventional: All leads have the same voltage with the II-lead as the reference
Amplitude: 0.5,1.0,1.5 and 2.0mV
The accuracy of Amplitude: ±2%
Rate: 15,20,25,30,40,60,80,100,120,140,160,180,200,220,240,260,280,300,320,340 and 350 BPM
artifacts: 50 or 60 Hz, muscle, baseline drift and breathing
The elevation of the ST segment: -0.8 mV to +0.8 mV, the step is 0.1 mV
The default: 80 BPM,1mV
Arrhythmia
Atrial: Atrial Fib(coarse), Atrial fib(fine), Atrial Flutter, Sinus Arrhythmia, Missed Beat, Atrial
Tachycardia, Nodal Rhythm, Supraventricular Tachycardia
Ventricular: Vent Fib (coarse), Vent Fib (fine), Premature Ventricular Contraction (PVC) 6/min,
PVCs 12/min, PVCs 24/min, Freqent Multifocal, Bigeminy, Trigeminy, Paired PVCs, Run 5
PVCs, Run 11 PVCs, Vent Tachycardia, Asystole
Premature Beat: Atrial Premature Atrial Contraction (PAC), Nodal PNC, Left Ventricular Focus
PVC1, Early Left Ventricular Focus PVC1, Right Ventricular Focus PVC1, Early Right
Ventricular Focus PVC1, R on T Left Ventricular Focus, R on T Right Ventricular Focus, Multifocal PVC
Conduction Defects: First Degree Heart Block, Second Degree Heart Block, Third Degree Heart Block, Right Bundle Branch Block, Left Bundle Branch Block
Performance Testing
amplitude0.5,1.0,1.5 and 2.0mV
The accuracy of Amplitude: ±2%
Pulse: 30 and 60 BPM, the width of pulse wave is 60 ms
Square: 0.125 and 2.0 Hz
Triangle: 2.0 and 2.5 Hz
Sine: .0.5, 5, 10, 40, 50, 60 and 100 Hz
Rate:15, 20, 25, 30, 40, 60, 80, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, 300, 320, 340 and 350 BPM
The accuracy of Rate: ±1%
Respiration
Respiration Rate: 0 (Apnea), 15, 20, 30, 40, 60, 80, 100, 120 and 150 R/MIIN
Respiration Impedance: 0.2, 0.5, 1.0 and 3.0 ¥Ø
Baseline Impedance:500,1000,1500 and 2000
Apnea: 12 s, 22 s, 32 s
simulator with adjustable heart rate/12-lead output
33 kinds of Arrhythmia waveforms
Pacemaker
Two IBP channels
Variety of respiratory impedance,apnea pauses (0 BrPM) of 12, 22 and 32 seconds.
Standard calibration
Export square,triangle and sine with adjustable frequencyIBP
The default: 00/00mmHg
Dynamic IBP
Channel 1: Artery 120/80mmHg
Radial 120/80mmHg
Left Ventricle 120/00mmHg
Right Ventricle 25/00mmHg
Channel 2: Artery 120/80mmHg
Radial 120/80mmHg
Left Ventricle 120/00mmHg
Right Atrium 15/10mmHg
Right Ventricle 25/00mmHg
Pulmonary Artery 25/10mmHg
Left Atrium 14/4mmHg
Static-Pressure Levels
Channel 1: 0, 80, 160, 240, 320 and 400mmHg
Channel 2: 0, 50, 100, 150, 200 and 250mmHg
ABOUT CONTEC MEDICAL SYSTEMS CO.,LTD
Contec Medical Systems focusing on research,manufacture and distribution of medical instruments,was founded in 1992 as a high-tech company. At present there are more than 1200 employees in our company.